Class I biological safety cabinets are primary barriers for the investigation of non-volatile materials or samples with unknown biological risk. After entrance of indoor air through front window aperture of the cabinet, air sweeps the work cabinet and surfaces in a non-turbulant way. The internal air exhausted from cabinet after passing through HEPA filter and protects the indoor environment from potentially dangerous particles. Since the Class I BSC does not protect sample/material, further aseptic or sterile analysis of working material is not suitable. 

Main Body & Design Parameters
  • Antibacterial Epoxy Powder paint Coated Steel
  • 300Openable hinged front panel for easy work-chamber cleaning
  • < 70angle sloped front panel with increased comfort & work-chamber vision
  • Electrical part protection inside at front panel isolated from working chamber
  • UV Resistant tempered glass
Microprocessor Control Unit & Display Parameters
  • One key touch on-off
  • Timer setting,
  • Inflow Air Velocity/Airflow Rate
  • Normal/Stand-by option button
  • Automated airflow compensation system
  • HEPA Filter Life (percent),
  • Total working Time,
  • UV and FL Lamp working times
  • Servicing records
  • Password protected technical service key
Audio-Visual Alarm Parameters
  • Audible and Visual Alarms for Air Flow/Rate 
  • System Ready/Not-Ready message/warning
  • Filter/Lamps replacement warning
  • Service need warning
  • Alarm cancel button
  • Unproper Sash window position warning (opt.)
Quality and Certificates
  • CE
  • ISO9001 Quality Management
  • EN 12469,
  • EN61010 Electrical and Electronic safety compatibility
  • Accredited test company validation guarantee (When installation made by Metis or by its certified technical personnel)
Optionals
  • High-performance HEPA Filter
  • ULPA Filter
  • Carbon Filter
  • Height adjustable support stand
  • Key-lock protection switch
  • Lux adjustable illumination
  • Active noise control system
  • Foot pedal control
  • Work chamber hanger apparatus
  • Heater & Microscope Assembly Base
  • Anti-vibration weighing balance table
  • Document / Equipment Cleaning Module
  • Motorized sash window
  • Sensor integrated Sash Window safety mechanism

Internal Dimentions (WxLxH) mm875x680x720
External Dimentions (WxLxH) mm995x833x1372
Support Stand (WxLxH) mm995x793x785
Support Stand Height From Ground855 mm
Inflow Air Flow Velocity (m/s)0.50 m/sn ± %5
Filters (EN 1822)Pre / Exhaust FilterF5 / Optional
Main FilterH14 HEPA, 0,3 µm particule %99.995 < filtration efficiency
Working TableStandart304 Stainless Steel
Optional316L Stainless Steel
Front Glass< 6 mm Lamine
Front Glass Aperture - Fixed (cm)46,5 cm
Motorized Front Glass (optional)Standart-Preset200cm ± 10mm
Maximum Height530cm ± 10mm
Noise LevelNormal Mod< 58 dB(A)
Eco (Stand by) Mode< 50 dB(A)
Working Area Illumination (Lux)850 – 1250 Lux
Power consumption (kW)Fan Filter Unit105 W
Illumination50 W
UV Lamp30 W
Electrical socket voltage and current220-240 VAC / 16A
Supply Voltage / Frequency /Power / Current230 VAC / 50 Hz / 2500 W / 16A
Pack/Palette Dimensions (WxLxH) mm1270x970x1770
Weight (kg), without support stand160 kg
Internal Dimentions (WxLxH) mm1180x680x720
External Dimentions (WxLxH) mm1300x833x1372
Support Stand (WxLxH) mm1300x793x785
Support Stand Height From Ground855 mm
Inflow Air Flow Velocity (m/s)0.50 m/sn ± %5
Filters (EN 1822)Pre / Exhaust FilterF5 / Optional
Main FilterH14 HEPA, 0,3 µm particule %99.995 < filtration efficiency
Working TableStandart304 Stainless Steel
Optional316L Stainless Steel
Front Glass< 6 mm Lamine
Front Glass Aperture - Fixed (cm)46,5 cm
Motorized Front Glass (optional)Standart-Preset200cm ± 10mm
Maximum Height530cm ± 10mm
Noise LevelNormal Mod< 58 dB(A)
Eco (Stand by) Mode< 50 dB(A)
Working Area Illumination (Lux)850 – 1250 Lux
Power consumption (kW)Fan Filter Unit200 W
Illumination50 W
UV Lamp30 W
Electrical socket voltage and current220-240 VAC / 16A
Supply Voltage / Frequency /Power / Current230 VAC / 50 Hz / 2500 W / 16A
Pack/Palette Dimensions (WxLxH) mm1575x970x1770
Weight (kg), without support stand190 kg
Internal Dimentions (WxLxH) mm1485x680x720
External Dimentions (WxLxH) mm1605x833x1372
Support Stand (WxLxH) mm1605x793x785
Support Stand Height From Ground855 mm
Inflow Air Flow Velocity (m/s)0.50 m/sn ± %5
Filters (EN 1822)Pre / Exhaust FilterF5 / Optional
Main FilterH14 HEPA, 0,3 µm particule %99.995 < filtration efficiency
Working TableStandart304 Stainless Steel
Optional316L Stainless Steel
Front Glass< 6 mm Lamine
Front Glass Aperture - Fixed (cm)46,5 cm
Motorized Front Glass (optional)Standart-Preset200cm ± 10mm
Maximum Height530cm ± 10mm
Noise LevelNormal Mod< 58 dB(A)
Eco (Stand by) Mode< 50 dB(A)
Working Area Illumination (Lux)850 – 1250 Lux
Power consumption (kW)Fan Filter Unit170 W
Illumination50 W
UV Lamp30 W
Electrical socket voltage and current220-240 VAC / 16A
Supply Voltage / Frequency /Power / Current230 VAC / 50 Hz / 2500 W / 16A
Ambalaj Boyutları (ExDxY) mm1880x970x1770
Weight (kg), without support stand220 kg
Internal Dimentions (WxLxH) mm1790x680x720
External Dimentions (WxLxH) mm1910x833x1372
Support Stand (WxLxH) mm1910x793x785
Support Stand Height From Ground855 mm
Inflow Air Flow Velocity (m/s)0.50 m/sn ± %5
Filters (EN 1822)Pre / Exhaust FilterF5 / Optional
Main FilterH14 HEPA, 0,3 µm particule %99.995 < filtration efficiency
Working TableStandart304 Stainless Steel
Optional316L Stainless Steel
Front Glass< 6 mm Lamine
Front Glass Aperture - Fixed (cm)46,5 cm
Motorized Front Glass (optional)Standart-Preset200cm ± 10mm
Maximum Height530cm ± 10mm
Noise LevelNormal Mod< 62 dB(A)
Eco (Stand by) Mode< 50 dB(A)
Working Area Illumination (Lux)850 – 1250 Lux
Power consumption (kW)Fan Filter Unit200 W
Illumination50 W
UV Lamp30 W
Electrical socket voltage and current220-240 VAC / 16A
Supply Voltage / Frequency /Power / Current230 VAC / 50 Hz / 2500 W / 16A
Ambalaj Boyutları (ExDxY) mm2185x970x1770
Weight (kg), without support stand280 kg

Metisafe Class I BSC Technical Drawing